Automatic Water Quality Analyzer for City Water (MWB4-72) is TOA-DKK Corporation's modern and potential product for continous monitoring water quality at faucet feed line or water receiving tank of a building.
*Measeasurements:
+ Turbidity
+ Color
+ Residual Chlorine
+ Electric Conductivity
+ pH
+ Temperature
+ Water Pressure
MWB4-72 also delivers high reliability, durability, and excellent maintainability.
Currently, GK Finechem Vietnam Co., Ltd is the official distributor of TOA-DKK for MWB4-72 in Vietnam.
Product name | Compact City Water Analyzer |
Model | : MWB4-72 |
Objects measured | : Three basic items (turbidity, color, and |
residual chlorine), electric conductivity, | |
pH, temperature, and pressure | |
Measurement | : Ability to switch between two ranges for |
range switching | both turbidity and color measurements. |
Display | : Color LCD touch panel |
Temperature | : 0 to 40°C for residual chlorine, EC, and |
compensation range | pH |
Response time | : 90% response within three minutes |
Operating power | : 100 to 240V AC ±10%, 50/60Hz |
Power consumption | : Approx. 40/55VA (100/240V AC), max. |
of approx. 83/108 VA (100/240V AC) | |
Transmission | : DC 4 to 20 mA, isolated |
output | ( Negative (-) side is common.) |
Load resistance | : 600Ω or less |
Contact switching | : Alarm 1; General alarms (measured value |
output signals | High-High/Low-Low limit alarm, light source |
error, residual chlorine motor error, sensor | |
error, and start-up mode error) | |
Alarm 2; General alarms (concentration | |
upper/lower limit alarm, water temperature | |
compensation error, and auto calibration error) | |
Maintenance; When in ST-BY mode | |
Event; During auto cleaning, during auto | |
calibration, and during problem diagnosis | |
(Contact capacity for all of the above; 24 | |
VDC 0.2A resistance load)) | |
Power cut off; The contact is closed when | |
a power failure occurs. | |
(Contact capacity; 30 VDC 0.2A | |
resistance load)) | |
Contact switching | : Cleaning command; When the closed |
input signals | contact receives, cell window cleaning |
starts (turbidity / color).Start cell window | |
cleaning (turbidity / color) | |
Calibration request; When the closed | |
contact receives, automatic zero | |
calibration starts (turbidity / color / residual | |
/ residual chlorine).Start automatic zero | |
calibration (turbidity / color / residual chlorine) | |
(Resistance load; 200Ωor less, Pulse | |
duration: 500mS or greater) | |
Communication | : RS232C interface or RS485 interface |
system | (isolated) |
Communication speed; 9600 BPS | |
Synchronous system; Start-stop | |
synchronization | |
Control system; Half-duplex | |
communication system | |
One line for communication (dedicated | |
cable or connector), one line for | |
maintenance (D-SUB connector) | |
Save functionality | : Data such as measurements can be |
transferred to a memory card.(Compact | |
Flash,CF) They can also be processed by | |
a computer. | |
One minute measurement values can be | |
stored in 3-month intervals, and one hour | |
values can be stored in 1-year intervals. | |
Sample water | : No suspension or stagnation. |
conditions | Quality; Ensures that the water quality |
(excluding the items below) satisfies the water | |
quality standards set by the Water Law or falls | |
within the measurement range of this unit. | |
Temperature; 0 to 40°C (no freezing) | |
Pressure; 0.05 to 0.75MPa | |
pH; 5.5 to 8.6 pH (maximum fluctuations; 1pH) | |
EC; 8 mS/m (80µS/cm) or greater | |
Flow rate; 50 to 100 mL/min | |
Sample | : 5m3/month or less (9m3/month, including |
consumption | a by-pass flow of 100 mL/min) |
Wetted part materials | : Polyurethane, PP, acrylic, stainless steel, |
FKM, etc. | |
Piping end connection | : Sample water inlet; Rc 1/4 |
Drain outlet; Rc 1/4 | |
Calibration solution inlet; Rc 1/4 | |
Air purge; Rc 1/4 | |
Installation | : Mounted on a wall or rack |
Wiring end connection | : Two water-proof connectors |
A power supply cable and I/O signal | |
cable (3-meter) are also included. | |
Ambient conditions | : 0 to 40°C (no freezing), 85% RH or less |
(no condensation) | |
Weight | : Approx. 11 kg |
Construction | : Indoor installation (IP43 equivalent) |
Case material | : Aluminum |
Color | : Light gray (Munsell 5PB 8/1 equivalent) |
Automatic calibration | : Zero calibration for turbidity, color, and |
residual chlorine. The internal timer and | |
an external contact signal are used to | |
start calibration. | |
(Zero water is prepared by filtering sample | |
water through a zero water filter.) | |
Calibration cycle setting; 0 to 24 hours | |
(freely specified) | |
Calibration time; Approx. 13 min. (fixed) | |
Transmission output hold time during | |
calibration; Calibration time approx. 13 | |
min. + 9 min. (fixed) | |
Automatic cleaning | : The internal timer and an external contact |
signal are used to start backwashing (by | |
draining the water) to clean the cell | |
window for turbidity and color measurements. | |
Cleaning cycle setting: 10, 15, 20, 30, or | |
60 min. | |
Transmission output hold time during | |
cleaning; Cleaning time approx. 2 min. + 1 | |
min. (fixed) | |
Beads cleaning of residual chlorine | |
electrode by self-rotation | |
Options | : Indoor self-standing frame (preassembled, piping pre-installed) |
Outdoor cubicle (temperature-controller | |
included) | |
Water sampling unit for abnormal time | |
Internal leakage detection unit |